Abstract
Objective Superior aortic valve hemodynamic performance can accelerate left ventricular mass regression and enhance survival and functional status after surgical aortic valve replacement. This can be achieved by rapid deployment aortic valve replacement using a subannular balloon-expandable stent frame, which functionally widens and reshapes the left ventricular outflow tract, to ensure a larger effective orifice area compared with conventional surgical valves. We report the intermediate-term follow-up data from a large series of patients enrolled in the Surgical Treatment of Aortic Stenosis With a Next Generation Surgical Aortic Valve (TRITON) trial.
Results The mean patient age was 75.7 ± 6.7 years (range, 45-93; 49.1% women). The mean aortic valve gradient significantly decreased from discharge to 3 years of follow-up. The mean effective orifice area remained stable from discharge to 3 years. At 1 year, the left ventricular mass index had decreased by 14% (P <.0001) and at 3 years by 16% (P <.0001) compared with at discharge. The prevalence of severe patient-prosthesis mismatch was 3% at discharge and remained stable during the follow-up period.
Conclusions In a large series of elderly patients with symptomatic severe aortic stenosis, rapid deployment aortic valve replacement using a subannular balloon-expandable stent frame demonstrated excellent hemodynamic performance and significant left ventricular mass regression. With continued follow-up, future studies will establish whether these favorable structural changes correlate with improvement in long-term survival and functional status.
| Original language | English |
|---|---|
| Pages (from-to) | 2854-2861 |
| Number of pages | 8 |
| Journal | Journal of Thoracic and Cardiovascular Surgery |
| Volume | 148 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - 01 Dec 2014 |
| Externally published | Yes |
Keywords
- Aged
- Aged, 80 and over
- Aortic Valve/diagnostic imaging
- Aortic Valve Stenosis/complications
- Bioprosthesis
- Europe
- Female
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation/adverse effects
- Hemodynamics
- Humans
- Hypertrophy, Left Ventricular/diagnosis
- Male
- Middle Aged
- Prospective Studies
- Prosthesis Design
- Time Factors
- Treatment Outcome
- Ultrasonography
- Ventricular Function, Left
- Ventricular Remodeling
ASJC Scopus subject areas
- Surgery
- Pulmonary and Respiratory Medicine
- Cardiology and Cardiovascular Medicine