Requirements for dose calculation on an active scanned proton beamline for small, shallow fields

B Knäusl, L Langgartner, M Stock, M Janson, K M Furutani, C J Beltran, D Georg, A F Resch

Research output: Journal article (peer-reviewed)Journal article

2 Citations (Scopus)

Abstract

INTRODUCTION: A growing interest in using proton pencil beam scanning in combination with collimators for the treatment of small, shallow targets, such as ocular melanoma or pre-clinical research emerged recently. This study aims at demonstrating that the dose of a synchrotron-based PBS system with a dedicated small, shallow field nozzle can be accurately predicted by a commercial treatment planning system (TPS) following appropriate tuning of both, nozzle and TPS.

MATERIALS: A removable extension to the clinical nozzle was developed to modify the beam shape passively. Five circular apertures with diameters between 5 to 34mm, mounted 72cm downstream of a range shifter were used. For each collimator treatment plans with spread-out Bragg peaks (SOBP) with a modulation of 3 to 30mm were measured and calculated with GATE/Geant4 and the research TPS RayStation (RS11B-R). The dose grid, multiple coulomb scattering and block discretization resolution were varied to find the optimal balance between accuracy and performance.

RESULTS: For SOBPs deeper than 10mm, the dose in the target agreed within 1% between RS11B-R, GATE/Geant4 and measurements for aperture diameters between 8 to 34mm, but deviated up to 5% for smaller apertures. A plastic taper was introduced reducing scatter contributions to the patient (from the pipe) and improving the dose calculation accuracy of the TPS to a 5% level in the entrance region for large apertures.

CONCLUSION: The commercial TPS and GATE/Geant4 can accurately calculate the dose for shallow, small proton fields using a collimator and pencil beam scanning.

Original languageEnglish
Article number102659
Pages (from-to)102659
JournalPhysica Medica
Volume113
Early online date18 Aug 2023
DOIs
Publication statusPublished - Sept 2023

Fingerprint

Dive into the research topics of 'Requirements for dose calculation on an active scanned proton beamline for small, shallow fields'. Together they form a unique fingerprint.

Cite this