TY - JOUR
T1 - Human Factors Evaluation of HeartMate 3 Left Ventricular Assist Device Peripherals
T2 - An Eye Tracking Supported Simulation Study
AU - Widhalm, Gregor
AU - Abart, Theodor
AU - Noeske, Moritz
AU - Kumer, Lisa
AU - Ebenberger, Katharina
AU - Atteneder, Clemens
AU - Berger, Angelika
AU - Laufer, Günther
AU - Wiedemann, Dominik
AU - Zimpfer, Daniel
AU - Schima, Heinrich
AU - Wagner, Michael
AU - Schlöglhofer, Thomas
N1 - Publisher Copyright:
© 2023, The Author(s).
PY - 2023/12
Y1 - 2023/12
N2 - Background: Despite recent design improvements, human factors issues continue to challenge left ventricular assist device (LVAD) therapy. The aim of this study was to evaluate user experience of former non-HeartMate 3 (HM3) LVAD patients post heart transplantation (HTX) and laypersons (LP) with HM3 LVAD peripherals in simulated everyday and emergency scenarios. Methods: This single center cohort study included untrained HTX and LP. Seven scenarios, including battery exchanges (without alarm, advisory alarm, dim light, consolidated bag), change of power supply, driveline dis-/reconnection and controller exchange were simulated. Subjects’ gaze behavior was recorded using eye tracking technology. Success rate, pump-off-time, duration to success (DTS), percental fixation duration per areas of interest and post-scenario-survey results were defined as outcome measures. Results: Thirty subjects completed 210 scenarios, initially solving 82.4% (HTX vs. LP, p = 1.00). Changing power supply revealed highest complexity (DTS = 251 ± 93s, p = 0.76): 26.7% succeeded at first attempt (p = 0.68), 56.7% at second attempt, with significantly more LP failing (p = 0.04), resulting in 10 hazards from driveline disconnections (pump-off-time 2-118s, p = 0.25). Comparison on initial success showed differences in fixation durations for seven areas of interest (p < 0.037). Decreasing DTS during battery exchanges (p < 0.001) indicate high learnability. Exchanging batteries within the bag took longer (median DTS = 75.0 (IQR = 45.0)s, p = 0.09), especially in elderly subjects (r = 0.61, p < 0.001). Subjects with less initial success were more afraid of making mistakes (p = 0.048). Conclusion: This eye tracking based human factors study provided insights into user experiences in handling HM3 peripherals. It highlights unintuitive and hazardous characteristics, providing guidance for future user-centered design of LVAD wearables.
AB - Background: Despite recent design improvements, human factors issues continue to challenge left ventricular assist device (LVAD) therapy. The aim of this study was to evaluate user experience of former non-HeartMate 3 (HM3) LVAD patients post heart transplantation (HTX) and laypersons (LP) with HM3 LVAD peripherals in simulated everyday and emergency scenarios. Methods: This single center cohort study included untrained HTX and LP. Seven scenarios, including battery exchanges (without alarm, advisory alarm, dim light, consolidated bag), change of power supply, driveline dis-/reconnection and controller exchange were simulated. Subjects’ gaze behavior was recorded using eye tracking technology. Success rate, pump-off-time, duration to success (DTS), percental fixation duration per areas of interest and post-scenario-survey results were defined as outcome measures. Results: Thirty subjects completed 210 scenarios, initially solving 82.4% (HTX vs. LP, p = 1.00). Changing power supply revealed highest complexity (DTS = 251 ± 93s, p = 0.76): 26.7% succeeded at first attempt (p = 0.68), 56.7% at second attempt, with significantly more LP failing (p = 0.04), resulting in 10 hazards from driveline disconnections (pump-off-time 2-118s, p = 0.25). Comparison on initial success showed differences in fixation durations for seven areas of interest (p < 0.037). Decreasing DTS during battery exchanges (p < 0.001) indicate high learnability. Exchanging batteries within the bag took longer (median DTS = 75.0 (IQR = 45.0)s, p = 0.09), especially in elderly subjects (r = 0.61, p < 0.001). Subjects with less initial success were more afraid of making mistakes (p = 0.048). Conclusion: This eye tracking based human factors study provided insights into user experiences in handling HM3 peripherals. It highlights unintuitive and hazardous characteristics, providing guidance for future user-centered design of LVAD wearables.
KW - Humans
KW - Aged
KW - Heart Failure/surgery
KW - Eye-Tracking Technology
KW - Cohort Studies
KW - Heart-Assist Devices
KW - Heart Transplantation
KW - Retrospective Studies
UR - https://www.scopus.com/pages/publications/85158035169
U2 - 10.1007/s10916-023-01950-3
DO - 10.1007/s10916-023-01950-3
M3 - Journal article
C2 - 37133553
SN - 0148-5598
VL - 47
SP - 58
JO - Journal of Medical Systems
JF - Journal of Medical Systems
IS - 1
M1 - 58
ER -