High-Intensity Transient Signals in the Outflow Graft and Thrombosis of a HeartWare Left Ventricular Assist Device

  • Kamen Dimitrov
  • , Julia Riebandt
  • , Thomas Haberl
  • , Dominik Wiedemann
  • , Paul Simon
  • , Günther Laufer
  • , Heinrich Schima
  • , Daniel Zimpfer

Research output: Journal article (peer-reviewed)Journal article

Abstract

We report the echocardiographic detection of microbubbles in the outflow prosthesis of a HeartWare HVAD left ventricular assist device using transthoracic echocardiography. In this case it was a specific sign associated with device thrombosis that resolved after successful treatment with systemic thrombolysis.

Original languageEnglish
Pages (from-to)e83-e85
JournalAnnals of Thoracic Surgery
Volume101
Issue number3
DOIs
Publication statusPublished - 01 Mar 2016
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Echocardiography/methods
  • Equipment Failure
  • Heart Diseases/diagnostic imaging
  • Heart Failure/surgery
  • Heart-Assist Devices/adverse effects
  • Humans
  • Male
  • Middle Aged
  • Thrombosis/diagnostic imaging

Fingerprint

Dive into the research topics of 'High-Intensity Transient Signals in the Outflow Graft and Thrombosis of a HeartWare Left Ventricular Assist Device'. Together they form a unique fingerprint.

Cite this