Abstract
The aim of this work is to present the first microdosimetric spectra measured with a miniaturised tissue-equivalent proportional counter in the clinical environment of the MedAustron ion-beam therapy facility. These spectra were gathered with a 62.4-MeV proton beam and have been compared with microdosimetric spectra measured in the 62-MeV clinical proton beam of the CATANA beam line. Monte Carlo simulations were performed using the Geant4 toolkit GATE and a fully commissioned clinical beam line model. Finally, similarities and discrepancies of the measured data to simulations based on a simple and complex detector geometry are discussed.
| Original language | English |
|---|---|
| Pages (from-to) | 1973-1978 |
| Number of pages | 6 |
| Journal | Radiation Protection Dosimetry |
| Volume | 199 |
| Issue number | 15-16 |
| DOIs | |
| Publication status | Published - 01 Oct 2023 |
| Externally published | Yes |
Keywords
- Protons
- Radiometry
- Proton Therapy
- Radiotherapy Dosage
- Monte Carlo Method
Fingerprint
Dive into the research topics of 'First microdosimetric measurements with a tissue-equivalent proportional counter at the MedAustron ion-beam therapy facility'. Together they form a unique fingerprint.Cite this