Abstract

Microdosemeters are frequently used today to specify the radiation quality in the framework of ion-beam therapy. The heterogeneity of the detector shapes and the materials limits the possibility of comparing directly spectra and mean lineal energies. A method was recently studied to convert the spectra obtained with unidirectional ion beams in slab detectors to those obtained with detectors of different in shape and material. The method is based on the observation that the lineal-energy spectra of slab detector, in a restricted energy interval, approximate the Linear Energy Transfer distributions at corresponding material and particle type and energies. In this study, the experimental spectra collected with a slab diamond detector are converted to the spectra that would be obtained using water detectors of spherical and cylindrical shapes.

OriginalspracheEnglisch
Seiten (von - bis)167-171
Seitenumfang5
FachzeitschriftRadiation Protection Dosimetry
Jahrgang183
Ausgabenummer1-2
DOIs
PublikationsstatusVeröffentlicht - 01 Mai 2019
Extern publiziertJa

ASJC Scopus Sachgebiete

  • Öffentliche Gesundheit, Umwelt- und Arbeitsmedizin
  • Strahlung
  • Radiologie- und Ultraschalltechnik
  • Radiologie, Nuklearmedizin und Bildgebung

Fingerprint

Untersuchen Sie die Forschungsthemen von „Microdosimetry of Clinical Ion Beams: Converting Spectra From Diamond Slab to Water of Different Shapes“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren