Influence of a fully magnetically levitated left ventricular assist device on functional interrogation of implantable cardioverter defibrillators

  • Christoph Schukro
  • , Thomas Schlöglhofer
  • , Cesar Khazen
  • , Michael Röhrich
  • , Günther Laufer
  • , Daniel Zimpfer
  • , Dominik Wiedemann

Publikation: Beitrag in Fachzeitschrift (peer-reviewed)Artikel in Fachzeitschrift

Abstract

BACKGROUND: Electromagnetic interference between left ventricular assist devices (LVAD) and the telemetry wand of implantable cardioverter-defibrillators (ICD) with impairment of ICD interrogation has previously been described in HVAD and HeartMate II devices. This is the first study showing the potential influence of the LVAD model HeartMate 3 (with the unique feature of fully magnetically levitated rotor with consistent wide blood-flow gaps) on functional interrogation of different ICD models.

METHODS AND RESULTS: Among 51 patients treated with a HeartMate 3 LVAD, 34 patients (66.7%) already had an ICD implanted prior to LVAD therapy. In this cohort, impairment of ICD interrogation was observed in five patients (14.7%) with five different device models. In patients with Biotronik ICD, stretching of the ipsilateral arm to increase the distance between both devices >10 cm was sufficient in one patients, whereas surgical contralateral repositioning was necessary in two patients; in one further patient no action could be taken, as he died early from embolic stroke. In the only patient with a MicroPort ICD, this issue was resolved by using a wireless telemetry. The distances between both devices showed no statistical significant correlation with an impaired interrogation, neither in the overall collective nor within the groups with the same manufacturer.

CONCLUSIONS: In patients with impaired ICD interrogation caused by electromagnetic interference between a HeartMate 3 LVAD and the ICD, the actions mentioned above have to be taken, to resolve this technical issue. Especially, a sufficient distance of at least 10 cm between both devices was crucial for avoiding this problem.

OriginalspracheEnglisch
Seiten (von - bis)914-918
Seitenumfang5
FachzeitschriftClinical Cardiology
Jahrgang42
Ausgabenummer10
DOIs
PublikationsstatusVeröffentlicht - 01 Okt. 2019