Abstract
Background: A relation between the left ventricular assist device inflow cannula (IC) malposition and pump thrombus has been reported. This study aimed to investigate if the pump position, derived from chest X-rays in HeartMate 3 (HM3) patients, correlates with neurological dysfunction (ND), ischemic stroke (IS), hemorrhagic stroke (HS) and survival. Methods: This analysis was performed on routinely acquired X-rays of 42 patients implanted with a HM3 between 2014 and 2017. Device position was quantified in patients with and without ND from frontal and lateral X-rays characterizing the IC and pump in relation to spine, diaphragm or horizontal line. The primary end-point was freedom from stroke and survival one-year after HM3 implantation stratified by pump position. Results: The analysis of X-rays, 33.5 (41.0) days postoperative, revealed a significant smaller IC angle of HM3 patients with ND versus no ND (0.1° ± 14.0° vs. 12.9° ± 10.1°, p = 0.005). Additionally, the IC angle in the frontal view, IS: 4.1 (20.9)° versus no IS: 13.8 (7.5)°, p = 0.004 was significantly smaller for HM3 patients with IS. Using receiver operating characteristics derived cut-off, IC angle <10° provided 75% sensitivity and 100% specificity (C-statistic = 0.85) for predicting IS. Stratified by IC angle, freedom from IS at 12 months was 100% (>10°) and 60% (<10°) respectively (p = 0.002). No significant differences were found in any end-point between patients with and without HS. One-year survival was significantly higher in patients with IC angle >10° versus <10° (100% vs. 71.8%, p = 0.012). Conclusions: IC malposition derived from standard chest X-rays serves as a risk factor for ND, IS and worse survival in HM3 patients.
| Originalsprache | Englisch |
|---|---|
| Seiten (von - bis) | 1149-1157 |
| Seitenumfang | 9 |
| Fachzeitschrift | Artificial Organs |
| Jahrgang | 46 |
| Ausgabenummer | 6 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - Juni 2022 |
| Extern publiziert | Ja |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gute Gesundheit und Wohlergehen
Fingerprint
Untersuchen Sie die Forschungsthemen von „Inflow cannula position as risk factor for stroke in patients with HeartMate 3 left ventricular assist devices“. Zusammen bilden sie einen einzigartigen Fingerprint.Dieses zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver