Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Human Factors Evaluation of HeartMate 3 Left Ventricular Assist Device Peripherals: An Eye Tracking Supported Simulation Study

  • Gregor Widhalm
  • , Theodor Abart
  • , Moritz Noeske
  • , Lisa Kumer
  • , Katharina Ebenberger
  • , Clemens Atteneder
  • , Angelika Berger
  • , Günther Laufer
  • , Dominik Wiedemann
  • , Daniel Zimpfer
  • , Heinrich Schima
  • , Michael Wagner
  • , Thomas Schlöglhofer

Publikation: Beitrag in Fachzeitschrift (peer-reviewed)Artikel in Fachzeitschrift

Abstract

Background: Despite recent design improvements, human factors issues continue to challenge left ventricular assist device (LVAD) therapy. The aim of this study was to evaluate user experience of former non-HeartMate 3 (HM3) LVAD patients post heart transplantation (HTX) and laypersons (LP) with HM3 LVAD peripherals in simulated everyday and emergency scenarios. Methods: This single center cohort study included untrained HTX and LP. Seven scenarios, including battery exchanges (without alarm, advisory alarm, dim light, consolidated bag), change of power supply, driveline dis-/reconnection and controller exchange were simulated. Subjects’ gaze behavior was recorded using eye tracking technology. Success rate, pump-off-time, duration to success (DTS), percental fixation duration per areas of interest and post-scenario-survey results were defined as outcome measures. Results: Thirty subjects completed 210 scenarios, initially solving 82.4% (HTX vs. LP, p = 1.00). Changing power supply revealed highest complexity (DTS = 251 ± 93s, p = 0.76): 26.7% succeeded at first attempt (p = 0.68), 56.7% at second attempt, with significantly more LP failing (p = 0.04), resulting in 10 hazards from driveline disconnections (pump-off-time 2-118s, p = 0.25). Comparison on initial success showed differences in fixation durations for seven areas of interest (p < 0.037). Decreasing DTS during battery exchanges (p < 0.001) indicate high learnability. Exchanging batteries within the bag took longer (median DTS = 75.0 (IQR = 45.0)s, p = 0.09), especially in elderly subjects (r = 0.61, p < 0.001). Subjects with less initial success were more afraid of making mistakes (p = 0.048). Conclusion: This eye tracking based human factors study provided insights into user experiences in handling HM3 peripherals. It highlights unintuitive and hazardous characteristics, providing guidance for future user-centered design of LVAD wearables.

OriginalspracheEnglisch
Aufsatznummer58
Seiten (von - bis)58
FachzeitschriftJournal of Medical Systems
Jahrgang47
Ausgabenummer1
DOIs
PublikationsstatusVeröffentlicht - Dez. 2023
Extern publiziertJa

Fingerprint

Untersuchen Sie die Forschungsthemen von „Human Factors Evaluation of HeartMate 3 Left Ventricular Assist Device Peripherals: An Eye Tracking Supported Simulation Study“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren