Helium pencil beam commissioning and beam modeling

  • Lukas Martin*
  • , Barbara Knäusl
  • , Peter Kuess
  • , Dietmar Georg
  • , Hugo Palmans
  • , Lorenz Wolf
  • , Nadia Gambino
  • , Fabio Farinon
  • , Andrej Prochazka
  • , Mansure Schafasand
  • , Lars Glimelius
  • , Walter Ikegami Andersson
  • , Daniel Simon Colomar
  • , Antonio Carlino
  • , Markus Stock
  • , Hermann Fuchs
  • *Korrespondierende/r Autor/-in für diese Arbeit

Publikation: Beitrag in Fachzeitschrift (peer-reviewed) › Artikel in Fachzeitschrift

Abstract

Background and Purpose: Helium ions combine reduced lateral scattering compared to protons and a lower fragmentation tail than carbon ions, enabling sharp dose gradients and improved normal tissue sparing. This study reports the commissioning of a scanned helium pencil beam line and validation of its corresponding beam model. Materials and Methods: Synchrotron-based helium ion beams were commissioned covering energies from 54.6 to 402.8MeV/u. Depth–dose curves were measured and absolute dose calibration was performed. Beam optics (spot size, position, and intraspill stability) were evaluated for various spill lengths. A beam model was implemented in the RayStation treatment planning system (TPS) and validated through 2D absolute dose measurements in homogeneous and heterogeneous phantoms and 3D measurements of cubic spread-out Bragg peak fields. Gamma-index analysis and Monte Carlo (MC) simulations with GATE/Geant4 were performed for benchmarking. Results: Measured ranges agreed with MC simulations within ±[jls-end-space/]0.3mm. Spot sizes decreased with energy, independently of the spill length. Spot positions remained within ±[jls-end-space/]0.5mm and intraspill variations were ≤[jls-end-space/]0.2mm (position) and ≤[jls-end-space/]5.4% (size). TPS-predicted doses agreed within 0.1%. For 3D validations in homogeneous phantoms, the dose differences were generally within 2%. Median gamma pass rates exceeded 90% for 3%/1.5mm and 95% for 5%/1.5mm. For the heterogeneous phantom, differences were within -3.2%. Conclusions: Stable scanned beam delivery with helium ions was established. Validation demonstrated strong agreement between measurements, TPS calculations, and MC simulations, supporting research and future clinical application.

OriginalspracheEnglisch
Aufsatznummer101046
Seiten (von - bis)101046
FachzeitschriftPhysics and Imaging in Radiation Oncology
Jahrgang40
DOIs
PublikationsstatusVeröffentlicht - 01 Juli 2026