HeartMate 3 Snoopy: Noninvasive cardiovascular diagnosis of patients with fully magnetically levitated blood pumps during echocardiographic speed ramp tests and Valsalva maneuvers

  • Thomas Schlöglhofer
  • , Christoph Gross
  • , Theodor Abart
  • , Anne-Kristin Schaefer
  • , Christiane Marko
  • , Michael Röhrich
  • , Gregor Widhalm
  • , Friedrich Kaufmann
  • , Ingo Weigel
  • , Hebe Al Asadi
  • , Barbara Karner
  • , Julia Riebandt
  • , Dominik Wiedemann
  • , Günther Laufer
  • , Heinrich Schima
  • , Daniel Zimpfer

Publikation: Beitrag in Fachzeitschrift (peer-reviewed)Artikel in Fachzeitschrift

Abstract

PURPOSE: The HeartMate 3 (HM3) left ventricular assist device (LVAD) has demonstrated excellent clinical outcomes; however, pump speed optimization is challenging with the available HM3 monitoring. Therefore, this study reports on clinical HM3 parameters collected with a noninvasive HM3 monitoring system (HM3 Snoopy) during echocardiographic speed ramp tests and Valsalva maneuvers.

METHODS: In this prospective, single-center study, the HM3 data communication between the controller and pump was recorded with a novel data acquisition system. Twelve pump parameters sampled every second (1 Hz) and clinical assessments (echocardiography, electrocardiogram (ECG), and blood pressure measurement) during speed ramp tests were analyzed using Pearson's correlation (r, median [IQR]). The cause for the occurrence of pulsatility index (PI)-events during ramp speed tests and valsalva maneuvers was investigated.

RESULTS: In 24 patients (age: 58.9 ± 8.8 years, body mass index: 28.1 ± 5.1 kg/m2, female: 20.8%), 35 speed ramp tests were performed with speed changes in the range of ±1000 rpm from a baseline speed of 5443 ± 244 rpm. Eight HM3 pump parameters from estimated flow, motor current, and LVAD speed together with blood pressure showed positive collinearities (r = 0.9 [0.1]). Negative collinearities were observed for pump flow pulsatility, pulsatility index, rotor noise, and left ventricular diameters (r = -0.8 [0.1]), whereas rotor displacement and heartrate showed absence of collinearities (r = -0.1 [0.08]).

CONCLUSIONS: In this study, the HM3 Snoopy was successfully used to acquire more parameters from the HM3 at a higher sampling rate. Analysis of HM3 per-second data provide additional clinical diagnostic information on heart-pump interactions and cause of PI-events.

OriginalspracheEnglisch
Seiten (von - bis)251-260
Seitenumfang10
FachzeitschriftJournal of Heart and Lung Transplantation
Jahrgang43
Ausgabenummer2
DOIs
PublikationsstatusVeröffentlicht - Feb. 2024
Extern publiziertJa

Fingerprint

Untersuchen Sie die Forschungsthemen von „HeartMate 3 Snoopy: Noninvasive cardiovascular diagnosis of patients with fully magnetically levitated blood pumps during echocardiographic speed ramp tests and Valsalva maneuvers“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren