Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Early Detection of Pump Thrombosis in Patients With Left Ventricular Assist Device

  • Joanna Grabska
  • , Thomas Schlöglhofer
  • , Christoph Gross
  • , Martin Maw
  • , Kamen Dimitrov
  • , Dominik Wiedemann
  • , Daniel Zimpfer
  • , Heinrich Schima
  • , Francesco Moscato

Publikation: Beitrag in Fachzeitschrift (peer-reviewed)Artikel in Fachzeitschrift

Abstract

Pump thrombosis (PT) is a serious adverse event in patients receiving left ventricular assist devices (LVAD). The study aims to determine whether pump parameters and clinical data may enable early detection of PT. This retrospective study included 88 patients who received an LVAD between 2012 and 2015 among which those with intra-PT were identified. In a propensity score-matched control group observation, time periods were matched with time before thrombosis. International normalized ratio (INR) time in therapeutic range (TTR) and lactate dehydrogenase (LDH) were analyzed for 60 days preceding PT. Furthermore, pump data (power, flow, and speed) in HeartWare ventricular assist devices (HVAD) patients were analyzed 7 days before PT using a mixed-design analysis of variance to investigate temporal changes in pump data. Pump thrombosis occurred in 15 patients (13 males, age 58 ± 10 years, 7 HeartMate II and 8 HVAD). International normalized ratio therapeutic range (2.0-3.0) and acetylsalicylic acid daily doses (100-200 mg) were similar for both groups, but patients with PT had lower TTR (36% vs. 65%; p = 0.025). No significant difference in LVAD power between groups was seen at baseline (p = 0.31), and power did not change in the control group over time (p > 0.99). Lactate dehydrogenase increased already 1 week prior PT and power from 4.4 ± 0.8 W at baseline to 4.9 ± 0.8 W (p = 0.007) 2 days before readmission and to 6.5 ± 1.8 W (p = 0.015) at readmission. Pump thrombosis is associated with a lower percentage of INR TTR and elevated LDH before the event. A better monitoring of pump parameters would enable PT detection already up to 2 days in advance.

OriginalspracheEnglisch
Seiten (von - bis)348-354
Seitenumfang7
FachzeitschriftASAIO Journal
Jahrgang66
Ausgabenummer4
DOIs
PublikationsstatusVeröffentlicht - 01 Apr. 2020
Extern publiziertJa

Fingerprint

Untersuchen Sie die Forschungsthemen von „Early Detection of Pump Thrombosis in Patients With Left Ventricular Assist Device“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren