Biomechanical characteristics of suture anchor implants for flexor digitorum profundus repair

Gabriel Halát*, Lukas Negrin, Thomas Koch, Jochen Erhart, Patrick Platzer, Stefan Hajdu, Johannes Streicher

*Korrespondierende:r Autor:in für diese Arbeit

Publikation: Beitrag in Fachzeitschrift (peer-reviewed)Artikel in Fachzeitschrift

10 Zitate (Scopus)

Abstract

Purpose To determine strength and failure characteristics of 2 suture anchors used to repair simulated flexor digitorum profundus avulsions during passive mobilization protocol simulation. Methods We simulated avulsion of the flexor digitorum profundus tendon in 30 distal phalanges from fresh-frozen human cadavers. Repair was performed with a 1.3 × 3.7 mm Micro-Mitek suture anchor (3-0 Orthocord suture) and a 2.2 × 4.0-mm Corkscrew suture anchor (2-0 FiberWire suture). All specimens were loaded cyclically from 2 to 15 N at 5 N/s for a total of 500 cycles. Samples were tested to failure at the completion of 500 cycles. Load at failure, load at first noteworthy displacement (> 2 mm), elongation of the system, gap formation at the tendon-bone interface, and the mechanism of failure were assessed. Results Suture failure at maximum load was the prevalent failure mechanism in both groups. No statistically significant difference in elongation of the tendon-suture complex was observed. The Corkscrew suture anchor showed a significantly superior performance in load to failure, load at first significant displacement, and gap formation at the tendon-bone interface. Conclusions The significantly higher load capacity at first displacement (> 2 mm) and the significance of a lower gap formation at the repair site seem to be the most relevant clinical parameters. Based on this concept, the Corkscrew anchor may be superior biomechanically to the Micro-Mitek when considering an early passive mobilization protocol. Clinical relevance The choice of an appropriate implant may influence the postoperative mobilization protocol and thereby improve currently reported success rates. Defining a biomechanically superior implant will provide an essential basis for further studies in flexor tendon repair research.

OriginalspracheEnglisch
Seiten (von - bis)256-261
Seitenumfang6
FachzeitschriftJournal of Hand Surgery
Jahrgang39
Ausgabenummer2
DOIs
PublikationsstatusVeröffentlicht - Feb. 2014
Extern publiziertJa

ASJC Scopus Sachgebiete

  • Chirurgie
  • Orthopädie und Sportmedizin

Fingerprint

Untersuchen Sie die Forschungsthemen von „Biomechanical characteristics of suture anchor implants for flexor digitorum profundus repair“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren